WEKO3
アイテム
Utility of a finger-mounted tissue oximeter with near-infrared spectroscopy to evaluate limb ischemia in patients with peripheral arterial disease
http://hdl.handle.net/10271/00003674
http://hdl.handle.net/10271/0000367420afb48a-0f12-46da-9b0d-8dd91a697074
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
|
Item type | 学位論文 / Thesis or Dissertation(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2019-12-10 | |||||
タイトル | ||||||
タイトル | Utility of a finger-mounted tissue oximeter with near-infrared spectroscopy to evaluate limb ischemia in patients with peripheral arterial disease | |||||
言語 | en | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | limb blood flow | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | near-infrared spectroscopy | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | peripheral arterial disease | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | regional hemoglobin oxygen saturation | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_db06 | |||||
資源タイプ | doctoral thesis | |||||
アクセス権 | ||||||
アクセス権 | open access | |||||
アクセス権URI | http://purl.org/coar/access_right/c_abf2 | |||||
その他のタイトル | ||||||
その他のタイトル | 末梢動脈疾患患者の下肢血流評価における近赤外分光法による指装着型オキシメータの有用性 | |||||
言語 | ja | |||||
著者 |
矢田, 達朗
× 矢田, 達朗 |
|||||
書誌情報 |
en : Annals of Vascular Diseases 巻 12, 号 1, p. 36-43, 発行日 2019-03-25 |
|||||
出版者 | ||||||
出版者 | Scientific Information Diseases | |||||
言語 | en | |||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | Objective: To investigate whether a finger-mounted tissue oximeter is useful in evaluating limb blood flow in patients with peripheral arterial disease (PAD). Materials and Methods: Seventy-two patients with PAD were included, and the ankle-brachial index (ABI), transcutaneous oxygen pressure (TcPO2), and skin perfusion pressure (SPP) were measured. The regional tissue oxygenation saturation (rSO2) was measured using a finger-mounted tissue oximeter at the ankle, dorsal foot, and each dorsal and plantar toe. Correlations between rSO2 and ABI and between TcPO2 and SPP were analyzed. The patients were divided into three groups: Fontaine IIa (F-IIa), IIb (F-IIb), and III and IV (F-III/IV) groups. The difference in rSO2 between each group was analyzed. Results: Significant correlations were observed between rSO2 and TcPO2 and between rSO2 and SPP. TcPO2 and SPP in the F-III/IV group were significantly lower than those in the F-IIa group. rSO2 in the F-IIb and F-III/IV groups was significantly lower than that in the F-IIa group. Conclusion: The measurement of rSO2 using fingermounted tissue oximetry is quick, simple, and painless. It can be used on any skin area and is useful to evaluate limb circulation in patients with PAD. |
|||||
言語 | en | |||||
学位名 | ||||||
言語 | ja | |||||
学位名 | 博士(医学) | |||||
学位の区分 | ||||||
内容記述タイプ | Other | |||||
内容記述 | doctoral | |||||
言語 | en | |||||
学位の分野 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 医学系研究科 | |||||
言語 | ja | |||||
学位授与機関 | ||||||
学位授与機関識別子Scheme | kakenhi | |||||
学位授与機関識別子 | 13802 | |||||
言語 | ja | |||||
学位授与機関名 | 浜松医科大学 | |||||
学位授与年月日 | ||||||
学位授与年月日 | 2019-03-13 | |||||
学位授与番号 | ||||||
学位授与番号 | 甲第804号 | |||||
ISSN | ||||||
収録物識別子タイプ | PISSN | |||||
収録物識別子 | 1881-641X | |||||
EISSN | ||||||
収録物識別子タイプ | EISSN | |||||
収録物識別子 | 1881-6428 | |||||
PubMed番号 | ||||||
識別子タイプ | PMID | |||||
関連識別子 | 30931055 | |||||
出版社DOI | ||||||
識別子タイプ | DOI | |||||
関連識別子 | https://doi.org/10.3400/avd.oa.18-00117 | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |