Rate-dependent elevation of the capture threshold after implantation of a leadless pacemaker |
|
言語 |
en |
言語 |
|
|
言語 |
eng |
キーワード |
|
|
主題 |
capture threshold |
キーワード |
|
|
主題 |
leadless pacemaker |
キーワード |
|
|
主題 |
phase 4 block |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
journal article |
アクセス権 |
|
|
アクセス権 |
open access |
|
アクセス権URI |
http://purl.org/coar/access_right/c_abf2 |
著者 |
Sano, Makoto
Takahashi, Naoyuki
Kaneko, Yutaro
Narumi, Taro
Naruse, Yoshihisa
Maekawa, Yuichiro
|
書誌情報 |
en : Pacing and Clinical Electrophysiology (PACE)
巻 47,
号 7,
p. 938-940,
発行日 2024-07-05
|
出版者 |
|
|
出版者 |
John Wiley and Sons |
|
言語 |
en |
権利 |
|
|
権利情報 |
This is the peer reviewed version of the following article: "Rate-dependent elevation of the capture threshold after implantation of a leadless pacemaker", 47(7), pp938-940, c2023, which has been published in final form at https://doi.org/10.1111/pace.14884. This article may be used for non-commercial purposes in accordance with Wiley Terms and Conditions for Use of Self-Archived Versions. This article may not be enhanced, enriched or otherwise transformed into a derivative work, without express permission from Wiley or by statutory rights under applicable legislation. Copyright notices must not be removed, obscured or modified. The article must be linked to Wiley’s version of record on Wiley Online Library and any embedding, framing or otherwise making available the article or pages thereof by third parties from platforms, services and websites other than Wiley Online Library must be prohibited. |
抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
The procedural success in the implantation of cardiac electric devices depends on both the implanted position and the electric performance. The capture threshold and the pacing output affect the estimated battery longevity. In a case with a high capture threshold, recapture and reimplantation of a leadless pacemaker are commonly recommended. We experienced a case with the rate-dependent elevation of the capture threshold following the implantation of a leadless pacemaker. The recognition of the rate-dependency of the capture threshold and the acceptable programming could avoid the unnecessary recapture and reimplantation of that, avoiding the increase of procedural risks. |
ISSN |
|
|
収録物識別子タイプ |
PISSN |
|
収録物識別子 |
0147-8389 |
EISSN |
|
|
収録物識別子タイプ |
EISSN |
|
収録物識別子 |
1540-8159 |
NII書誌ID |
|
|
収録物識別子タイプ |
NCID |
|
収録物識別子 |
AA00766720 |
PubMed番号 |
|
|
関連タイプ |
isVersionOf |
|
|
識別子タイプ |
PMID |
|
|
関連識別子 |
38010832 |
出版社DOI |
|
|
関連タイプ |
isVersionOf |
|
|
識別子タイプ |
DOI |
|
|
関連識別子 |
https://doi.org/10.1111/pace.14884 |
著者版フラグ |
|
|
出版タイプ |
AM |
|
出版タイプResource |
http://purl.org/coar/version/c_ab4af688f83e57aa |